Fluke 718Ex Intrinsically Safe Pressure Calibrator

Model#:Fluke 718Ex 30GPart#:2096310UPC#:0 95969 24164 9
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 185.47ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • Fluke 718Ex Intrinsically Safe Pressure Calibrator
  • Fluke 718Ex Intrinsically Safe Pressure Calibrator

Key features

  • ATEX II 1G Ex ia IIC T4 Compliant.
  • Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
  • 30, 100, and 300 psi ranges (2, 7, and 20 bar)
  • Pressure measurement to 0.05% of full span, using an internal pressure sensor.
  • Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
  • Wide range of selectable measurement units for pressure.
  • Current measurement with 0.02% accuracy and 0.001 mA resolution.
  • Pressure switch test function.
  • Min-Max Hold functions.
  • Compact size and weight.
  • Simple, push-button user interface.
  • Rugged and reliable, ideal for field use.

Sale ends soon.

Shop now!

DAYS
HR
MIN
SEC
In stock
$5,149.99
FREE Shipping
30-day Returns

Product overview: Fluke 718Ex Intrinsically Safe Pressure Calibrator

Pump Up the Pressure

The Fluke 718Ex Intrinsically Safe Pressure Calibrator is a powerful intrinsically safe, self contained pressure calibrator for usage in explosion endangered areas.

Related blog posts:

Calibrating in potentially explosive atmospheres

Specifications: Fluke 718Ex Intrinsically Safe Pressure Calibrator

Pressure Measurement Accuracy
Fluke 718Ex 30GRange-830 mbar to 2 bar (internal sensor)
Over pressure: 3x FS
Resolution0.0001 bar
Accuracy0.05% FS
FunctionsZero, Min, Max, Hold, Damp
Fluke 718Ex 100GRange-830 mbar to 2 bar (internal sensor)
Over pressure: 2x FS
Resolution0.001 bar
Accuracy0.05% FS
FunctionsZero, Min, Max, Hold, Damp
Fluke 718Ex 300GRange-830 mbar to 20 bar (internal sensor)
Over pressure: 375 psi, 25 bar
Resolution0.0001 bar
Accuracy0.05% FS
FunctionsZero, Min, Max, Hold, Damp
Fluke 718Ex seriesRangeVia pressure module connector
29 pressure modules, 2.5 mbar to 700 bar
Over pressure per pressure module specs
ResolutionPer pressure module specs
AccuracyTo 0.025% of full span, per pressure module specs
FunctionsZero, Min, Max, Hold, Damp
Source pressure, built in pump range-820 mbar to full scale
For use with non-corrosive gasses.
Supported Pressure Units psi, inH2O (4°C), inH2O (20°C), kPa, cmH2O (4°C), cmH2O (20°C), bar, mbar, kg/cm², mmHg, inHg
Media compatibility per pressure module specs
Technical Data
Measure accuracyRange0-24 mA
Resolution0.001 mA
Accuracy0.02% rdg ±2 counts
DisplayLCD, 5 digit pressure and current simultaneous
Environmental Specifications
Operating temperature-10°C to 55°C
Storage temperature-40°C to 60°C
Humidity (without condensation)95% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
35% (50°C to 55°C)
Operating altitude3,000 m max
Safety Specifications
SafetyATEX II 1G Ex ia IIC T4 compliant
CSA Certification I.S. Class I
Div 1 Groups A-D , T4
Mechanical and General Specifications
Size (L x W x H)216 x 94 x 66 mm (with holster)
WeightApprox. 992 g (with holster)
WarrantyThree years, one year for pump
Shock1 meter drop test
VibrationRandom, 2 G, 5-500 Hz
Battery typeOne 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V
Battery life4 to 20 hours, depending on functions used

Models: Fluke 718Ex Intrinsically Safe Pressure Calibrator

Fluke 718Ex 30G
Fluke 718Ex 30G Intrinsically Safe Pressure Calibrator, 30 psi
Part#: 2096310
UPC#: 0 95969 24164 9

Includes:

  • Protective red holster
  • TL75 test leads
  • AC72 test clips
  • CD users manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
  • Fluke 718Ex CCD content control drawing
  • NIST- traceable calibration report and dat
Fluke 718EX 100G
Fluke 718Ex 100G Intrinsically Safe Pressure Calibrator, 100 psi
Part#: 2096322
UPC#: 0 95969 24165 6

Includes:

  • Protective red holster
  • TL75 test leads
  • AC72 test clips
  • CD users manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
  • Fluke 718Ex CCD content control drawing
  • NIST- traceable calibration report and data
Fluke 718EX 300G
Fluke 718Ex 300G Intrinsically Safe Pressure Calibrator, 300 psi
Part#: 3023328
UPC#: 0 95969 41386 2

Includes:

  • Protective red holster
  • TL75 test leads
  • AC72 test clips
  • CD users manuals (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese)
  • Fluke 718Ex CCD content control drawing
  • NIST- traceable calibration report and dat

We bought a new 718ex-300G pressure calibrator new, the calibration certificate due date is that 1 year from when we receive it? let me know thank you much!

Asked by: Jacob R
The certification due date will be 1 year from the issue date on the calibration certificate.
Answered by: FlukeCorp
Date published: 2023-09-15
  • y_2024, m_11, d_19, h_8CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_US, sid_3840954105908, prod, sort_[SortEntry(order=LAST_APPROVED_ANSWER_SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_fluke
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 195.91ms
  • QUESTIONS, PRODUCT
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 99.06ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.