ADCS-601 Air Data Calibration Standard

  • No image available
Wycofane
Ten produkt został wycofany z produkcji lub przeklasyfikowany

Omówienie produktu: ADCS-601 Air Data Calibration Standard

The ADCS-601 system is made up of two automated pressure standards with overlapping ranges to cover the full pressure range. An automated PG7601 piston gauge covers the range of 15 to 380 kPa (4.5 to 120 inHg) (see Section 2.2.1). The PG7601 range is extended down to 1 kPa (0.3 inHg) by an FPG8601 force balanced piston gauge with automated pressure control (see Section 2.2.2). The system also includes vacuum pumps and vacuum measurement instruments to support absolute mode operation. An RPM4 Reference Pressure Monitor is included to assist in performing a comparison of the FPG8601 and PG7601 pressure standards.

Physically, the ADCS-601 system is divided into two major components. The Reference Bench is a laboratory table onto which are installed the PG7601 and FPG8601 pressure standards, pneumatic control switching hardware and the reference vacuum pumping systems. The Control Cabinet contains the PPC3 automated pressure controllers, RPM4 Reference Pressure Monitor and Gas Supply Panel. The system controller (personal computer) is connected to the system by a USB interface hub.

The ADCS-601 system is controlled by ADCS Tools software running on a personal computer. ADCS Tools manages the entire system, communicating with other software programs and system instruments. ADCS Tools also provides the overall system interface through the personal computers display, keyboard and pointing device.

Reference bench

The ADCS-601 Reference Bench is a wooden topped laboratory work bench onto which are installed the PG7601 and FPG8601 pressure standards. Also included on the bench is a pneumatic interconnections box (see Figure 8), the system connections to the device under test (DUT) and the pumping system for the reference vacuum against which absolute pressure is defined. The vacuum pumps may be purchased separately.

Control cabinet

The ADCS-601 Control Cabinet is a 19-inch rack-mount enclosure into which are mounted the PPC3 pressure controllers that provide automated pressure control for the PG7601 and FPG8601 pressure standards and an RPM4 Reference Pressure Monitor used to perform the validation comparison of the PG7601 and FPG8601. Also included in the control cabinet are electrical power connections and a gas supply panel which sets and indicates gas pressure supplies to various parts of the system.

Gas supply panel

The Gas Supply Panel is in the ADCS-601 Control. It is accessed by a door in the lower part of the front of the Control Cabinet. The Gas Supply Panel is used to set and indicate gas pressure supplies to various ADCS-601 systems and components.

Specyfikacje: ADCS-601 Air Data Calibration Standard

Specifications
Power requirementsInstruments (AC1)
100 to 240 VAC, 50/60 Hz, 240 W max. consumption
Fuse: 250VT 10A

Reference vacuum pumps (AC2)
110 V version: 100 – 120V, 50/60 Hz, 1100 W max. consumption
Fuse: 250VT 10A
220 V version: 200 – 240V, 50/60 Hz, 1100 W max. consumption
Fuse: 250VT 10A

Utility pump (AC3)
110 V version: 100 – 120V, 50/60 Hz, 800 W max. consumption
Fuse: 250VT 10A
220 V version: 200 – 240V, 50/60 Hz, 800 W max. consumption
Fuse: 250VT 10A
Pressure supply requirementsDrive air
0.7 to 1 MPa (100 to 150 psi)

Test gas
Nitrogen or air, 0.45 to 1 MPa (65 to 150 psi), clean and dry (instrument grade minimum, high purity preferred)
Operating temperature range20 ºC to 26 ºC
WeightReference bench
173 kg (380 lb) approx. (with turbo pump and controller installed)

Control cabinet
105 kg (230 lb) approx.

Utility and roughing vacuum pumps
19 kg (42 lb) each
DimensionsReference bench
145 cm H x 122 cm W x 76 cm D (57 in. x 48 in. x 30 in.)

Control cabinet
124 cm H x 56 cm W x 66 cm D (49 in. x 22 in. x 26 in.)
Pressure rangesOverall
1 to 380 kPa (0.3 to 120 inHg)

PG7601
10 to 380 kPa (3 to 120 inHg)

FPG8601
1 to 15 kPa (0.2 to 4.5 inHg)
Operating mediumNitrogen or air
Pressure connectionsTEST(+) port
Quick connector (1/4 in. NUPRO® SS-QC4-B1-400)

TEST(-) port
Quick connector (1/4 in. NUPRO® SS-QC4-B1-400)
CD conformanceAvailable, must be specified

PG7601 Piston Gauge Specifications

Ambient and piston-cylinder condition measurements
TemperatureRange
Ambient: 0 °C to 40 °C
Piston-cylinder module: 0 °C to 40 °C
Resolution
Ambient: 0.1
Piston-cylinder module: 0.01
Measurement uncertainty
Ambient: ± 1
Piston-cylinder module: ± 0.1
Barometric pressureRange
70 kPa to 110 kPa
Resolution
10 Pa
Measurement uncertainty
± 140 Pa
Relative humidityRange
5 to 95% RH
Resolution
1% RH
Measurement uncertainty
± 10% RH
Piston positionRange
± 4.5 mm
Resolution
0.1 Mm
Measurement uncertainty
± 0.2 mm
Piston rotation (rate and deceleration)Range
2 to 99 rpm
Resolution
1 rpm
Reference vacuumWith on-board gauge
Range
0 to 20 Pa
Resolution
0.01 Pa
Measurement uncertainty
± 0.1 Pa or 10% of reading, whichever is greater

With external capacitance diaphragm gauge
Range
0 to 13.3 Pa (100 mTorr)
Resolution
0.01 Pa (0.01 mTorr)
Measurement uncertainty
± 0.05 Pa (0.38 mTorr)
Piston-cylinder module: PC-7100/7600-10
Nominal pressure to mass10 kPa/kg
OperationGas operated, gas lubricated
Piston materialTungsten carbide
Cylinder materialTungsten carbide
Nominal diameter35 mm
Nominal area1,000 mm2
Mounting systemPositive free deformation
Typical drop rate0.1 mm/min
Piston assembly nominal mass0.5 kg
Mass set: masses > 50g
Material304L non-magnetic stainless steel
FinishElectropolished
Uncertainty in measured values± 5 ppm or 1 mg, whichever is greater

FPG8601 General Specifications

Ambient and piston-cylinder condition measurements
TemperatureLubrication gas
Range: 0 °C to 40 °C
Resolution: 0.1 °C
Accuracy: ± 0.2 °C

Piston-cylinder
Range: 0 °C to 40 °C
Resolution: 0.01 °C
Accuracy: ± 0.1 °C
Relative humidityRange: 5 to 95% RH
Resolution: 1% RH
Accuracy: ± 10 %RH
VacuumRange: 13.3 Pa (100 mTorr)
Resolution: 0.001 Pa (0.01 mTorr)
Accuracy: ± 0.06 Pa (0.45 mTorr)
Ambient and lubrication pressure sensorRange: 200 kPa
Resolution: 1 Pa
Accuracy: ± 0.1 kPa

Pressure measurement and control specifications

10 to 380 KPA, PG7601 and PPC3-700K
Sensitivity¹0.01 Pa + 0.5 ppm
Precision²± (0.01 Pa + 2 ppm)
Measurement uncertainty³see the piston-cylinder calibration report and current revision of Technical Note 7920TN01.
Control set point increments1 kPa
Stability of set pressure± 1 Pa
Typical control set point time60 to 180 seconds

¹ Sensitivity: The smallest variation in input detectable in output.
² Precision: Combined linearity, hysteresis, repeatability.
³ Measurement uncertainty: Maximum deviation of the indicated pressure from the true value of applied pressure including all sources of uncertainty combined and expanded (k=2) following the ISO "Guide to the Expression of Uncertainty in Measurement."

1 to 15 KPA, FPG8601 and PPC3-100K
Resolution¹0.01 Pa
Precision²± (2 ppm + 50 mPa)
Measurement uncertainty³± 0.008% of reading or 0.3 Pa (0.0001 inHg) , whichever is larger
Control set point increments1 Pa
Stability of set pressure± 1 Pa (0.0003 inHg)
Typical control set point time60 to 180 seconds

¹ Resolution: Minimum display increment.
² Precision: Combined linearity, hysteresis, repeatability.
³ Measurement uncertainty: Maximum deviation of the indicated pressure from the true value of applied pressure including all sources of uncertainty combined and expanded (k=2) following the ISO “Guide to the Expression of Uncertainty in Measurement." Note: The uncertainty assigned to the FPG8601 used in an ADCS-601 system is expanded relative to the standard FPG8601 instrument. to allow validation by the PG7601 and an indefinite recalibration interval. See Section 7.2.3 and Technical Note 2090TN05.

Modele: ADCS-601 Air Data Calibration Standard

ADCS-601
Air Data Calibration Standard