5500A Multi-Product Calibrator
Non più disponibile
Questo prodotto non è più disponibile o è stato riclassificato
Fluke suggerisce:5502A Multi-Product Calibrator
Descrizione generale del prodotto: 5500A Multi-Product Calibrator
Dati tecnici: 5500A Multi-Product Calibrator
Technical Specifications | |
Voltage DC | Range: 0 to ±1020V Best 1 Year Specification: ±50 ppm of setting |
Current DC | Range: 0 to ±11A Best 1 Year Specification: ±0.01% of setting |
Resistance | Range: 0 to 3.29.999 MΩ Best 1 Year Specification: ±0.009% of setting |
Voltage AC | Range: 1 mV to 1020V 10 Hz to 500 kHz, Sine Best 1 Year Specification: ±0.03% of setting |
Current AC | Range: 29 m A to 11A 10 Hz to 10 kHz, Sine Best 1 Year Specification: ±0.06% of setting |
Capacitance | Range: 0.33 nF to 1.1 mF Best 1 Year Specification: ±0.25% of setting |
Thermocouple Source and Measure | Range: -250°C to +2316°C (11 types) Best 1 Year Specification: ± 0.14°C of setting |
RTD Source | Range: -200°C to +630°C (8 types) Best 1 Year Specification: ±0.03°C of setting |
DC Power | Range: 109 mW to 11 kW Best 1 Year Specification: ± 0.08% of setting |
AC Power | Range: 109 mW to 11 kW 45-65 Hz, PF=1 Best 1 Year Specification: ± 0.15% of setting |
Phase | Range: 0 to ± 179.99° Best 1 Year Specification: ±0.15° |
Frequency | Range: 0.01 Hz to 2.0 MHz Best 1 Year Specification: ±25 ppm of setting |
General Specifications | |
Warm up Time | Twice the time since last warmed up, to a maximum of 30 minutes |
Settling Time | Less than 5 seconds for all functions and ranges except as noted |
Standard Interfaces | IEEE-488 (GPIB), RS-232, 5725A (5500A only) |
Temperature Performance | Operating: 0°C to 50°C Calibration: (tcal): 15°C to 35°C Storage: -20°C to 70°C |
Temperature Coefficient | Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C |
Relative Humidity | Operating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C Storage: < 95%, non-condensing |
Altitude | Operating: 3,050m (10,000 ft) maximum Non-operating: 12,200m (40,000 ft) maximum |
Safety | Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92 |
Analog Low Isolation | 20V |
EMC | Designed to comply with FCC Rules Part 15 |
Line Power | Line Voltage (selectable): 100V, 120V, 220V, 240V Line Frequency: 47 to 63 Hz Line Voltage Variation: : ± 10% about line voltage setting |
Power Consumption | 300 VA |
Calibration Documentation | 17025 accredited report of calibration included |
Dimensions | Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit Width: 43.2 cm (17 in), standard rack width Depth: 47.3 cm (18.6 in) overall |
Weight (without options) | 20 kg (44 lbs.) |
Absolute Uncertainty Definition | 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated. |
Specification Confidence Interval | > 99% |
Modelli: 5500A Multi-Product Calibrator
5500A
Multi-Product Calibrator
Manuali + risorse: 5500A Multi-Product Calibrator
- 5500A | Manuale di Servizio
- 5500A | Supplemento al manuale di assistenza
- 5500A | Manuale programmatori
- 5500A | Manuale per gli operatori
- 5500A | Guida di riferimento
- 5500A | Manuale per gli operatori
- 5500A | Supplemento al manuale per gli operatori
- 5500A | Informazioni introduttive
- 5500A | Dichiarazione di volatilità della memoria
- 5500A/COIL | Dichiarazione di volatilità della memoria
- 5500A/COIL | Foglio di istruzioni
- 550XA/525A/Leads | Foglio di istruzioni
- Shielded Calibration Test Leads Test Leads | Foglio di istruzioni
- Y5537A | Foglio di istruzioni