Fluke 787 ProcessMeter Digital Multimeter

  • Fluke 787 ProcessMeter™
  • Fluke 787 ProcessMeter™
Non più disponibile

Caratteristiche principali

  • DMM and a loop calibrator in one rugged, handheld tool
  • Provides a precision 1000 V, 440 mA True-RMS digital multimeter designed to meet 1000 volt IEC 1010 CAT III standards
  • Features a clear LCD with backlight; 4,000 counts (30,000 counts for DC current)
  • Performs frequency measurements to 20 kHz
Questo prodotto non è più disponibile o è stato riclassificato

Descrizione generale del prodotto: Fluke 787 ProcessMeter Digital Multimeter

Fluke 787 ProcessMeter doubles your capabilities but not your cost

The Fluke 787 ProcessMeter combines a DMM and a Loop Calibrator in one rugged, handheld tool, for about what you would expect to pay for a loop calibrator alone. It's a maintenance and calibration tool that will be at home in every instrumentation technician's tool box. Based on the trusted Fluke 87 DMM, the 787 adds the ability to measure, source, and simulate dc loop current with 1 microamp resolution and 0.05% accuracy. It puts a complete solution for troubleshooting and calibrating current loop applications in the palm of your hand so you can get more information, faster and easier.

So if you're tired of having to track down a calibrator, lugging around two tools, and switching back and forth, get the convenient solution for doubling your power. The 787 ProcessMeter from Fluke.

Other useful features:

  • Provides Min/Max/Average/Hold/Relative modes, and diode test and continuity beeper
  • Displays simultaneous mA and % of scale readout
  • Provides 20 mA DC current source/loop calibrator/simulator
  • Manual step (100%, 25%, Coarse, Fine) plus auto step and auto ramp
  • Externally accessible battery for easy battery changes
  • 1000 V overload protection on V, ohms, frequency
  • 150 V overload protection on mA, backed up by 440 mA 1000V fuse

Want a built in 24-Volt loop supply and 250 Ω HART resistor? Consider the Fluke 789.

Dati tecnici: Fluke 787 ProcessMeter Digital Multimeter

Specifications
Current measurementRange0-1 A
Resolution1 mA
Accuracy0.2% + 2 LSD
Range0-30 mA
Resolution0.001 mA
Accuracy0.05% + 2 LSD
Current sourcingRange0-20 mA or 4-20 mA
Resolution0.05% of span
Drive Capability500 Ω @ 24 mA
FrequencyTo 19.999 kHz, 0.005% + 1 LSD
ContinuityBeeps for resistance < 100 ohms
Voltage measurementRange0-1000 V AC or DC
Resolution0.1 mV to 1.0 V
Accuracy0.1% Rdg + 1 LSD (VDC)
Resistance measurementTo 40 MΩ, 0.2% + 1 LSD
Diode test2.4 V shows diode voltage drop
Environmental Specifications
Operating temperature-20°C to 55°C
Storage temperature-40°C to 60°C
Humidity (without condensation)95% up to 30°C
75% up to 40°C
45% up to 50°C
35% up to 55°C
Temperature coefficient0.05 x (specified accuracy) per °C (for temperatures < 18°C or > 28°C)
Safety Specifications
Electrical safetyDesigned in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92
Overvoltage categoryIEC 1010-1 CAT III
Maximum voltage1000 V rms
Mechanical and General Specifications
Size32 x 87 x 187 mm
Weight369 g
BatteriesSingle 9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
Warranty3 years
Battery life50 hours typical (measurement), 12 hours typical (sourcing 12 mA)
Shock and vibrationRandom, 2 g, 5-500 Hz
1 meter drop test
Display current and % of spanYes
Auto step, auto rampYes

Modelli: Fluke 787 ProcessMeter Digital Multimeter

Fluke 787

Tutti i modelli includono:


Includes:

  • AC70A Alligator Clips
  • C81Y Protective Yellow Holster with Flex-Stand™
  • One 9V Alkaline Battery
  • Quick Reference Guide
  • TL75 Safety-Designed Test lead Set