A40B Series Precision DC and AC Current Shunts

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 21.62ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;

Key features

Read more

Product overview: A40B Series Precision DC and AC Current Shunts

Priority Gold Instrument CarePlan

Make sure to ask your account manager about the Fluke Calibration Priority Gold CarePlan

A Fluke Calibration Priority Gold CarePlan puts you in control of your downtime and in control of your business, with calibration services and other offerings.

Video Player is loading.
Current Time 0:00
Duration 0:00
Loaded: 0%
Stream Type LIVE
Remaining Time 0:00
 
1x
    • Chapters
    • descriptions off, selected
    • subtitles off, selected

      Learn more

      • Simplifies calibration/verification of precision calibrators and current sources
      • Shunts sized for currents from 1mA to 100A
      • Usable from dc to 100 kHz
      • 14 individual dc and ac current shunts with a 1, 2, 5 sequence over 6 decades of current
      • Simple direct measurements, making ac/dc transfers unnecessary
      • Stability typically better then ± 5.0 µΩ/Ω for one year
      • Typical angular accuracy of better than ± 0.003º at 1 kHz

      Application

      The wide range of shunt values permit measuring from 0.1 mA to 100 Amps. The physical construction and the components used in the current shunts ensure that the frequency response is very flat (amplitude displacement error relative to dc resistance). Also the phase displacement at 100 kHz is small enough to be neglected in all but the highest accuracy measurements. Combined with excellent dc resistance stability, the shunts can be used to directly measure current through their full bandwidth. This simplifies precision ac current measurements as the complex ac/dc transfer measurement process will no longer be necessary for many precision current measurement applications. These characteristics make the A40B shunts ideal for traditional precision current applications such as calibrator verification. Also, the wide current range allows verification of high current transconductance amplifiers. The low phase shift error is critical for measurement of non-sinusoidal wave shapes, as found in power quality or sampling digital wattmeter measurement applications.

      Technical overview

      The shunts come in four different sizes. The full current shunt set includes:

      • four fully enclosed dc and ac current shunts: 1 mA, 10 mA, 20 mA, 50 mA
      • five small size radial dc and ac current shunts: 100 mA, 200 mA, 500 mA, 1 A, 2 A
      • three medium size radial dc and ac current shunts: 5A, 10 A, 20 A
      • two large size radial dc and ac current shunts: 50 A, 100 A

      Voltage output is nominally 0.8 volts for the nominal rated current input. The output is measured by a voltmeter or detector, such as precision voltmeters, ac measurement standards, ac/dc transfer standards, or thermal voltage converters, fitting them to a wide variety of metrology applications.

      The radial style dc and ac current shunt design gives high performance with minimum external magnetic fields. Plus, the open nature of the dc and ac current shunt physical design maximizes air flow so the shunts have a minimal power coefficient effects. This permits each shunt to be used over a wide range of currents with stable resistance characteristics.

      The dc and ac current shunt resistance is designed to minimize interaction with the detector instrumentation. The 100 Amp shunt resistance is 8 milliohms, with shunt resistance increasing to 80 ohms for the 10 mA shunt. Additionally, the 1mA shunt includes an internal battery operated buffer amplifier. It drives the output voltage so the 800 ohm shunt resistance has minimal interaction with the measurement device.

      Reports of calibration

      The A40B series of precision DC and AC Current Shunts come with an ISO 17025 Accredited Certificate of Calibration.

      Specifications: A40B Series Precision DC and AC Current Shunts

      Basic Electrical Specification Summary

      Shunt
      Nominal
      Current
      Nominal
      Resistance
      (Ohms)
      Basic Specification ±μA/A, TCal ±1 °C, ≤50 % RH
      DC1 kHz10 kHz30 kHz100 kHz
      1 mA80020557575150
      10 mA802026262626
      20 mA402026262626
      50 mA162023232323
      100 mA82024242424
      200 mA42026262626
      500 mA1.62127272728
      1 A0.82127282831
      2 A0.42127303048
      5 A0.162131324071
      10 A0.082637606192
      20 A0.0426435270113
      50 A0.01632558081144
      100 A0.00835659098174

      Maximum AC-DC Difference

      Shunt Nominal Current:Maximum AC-DC Difference (± ppm)1,2
       1 kHz100 kHz
      1 mA353150
      10 mA2040
      20 mA1830
      50 mA1316
      100 mA1427
      200 mA1728
      500 mA1721
      1 A1723
      2 A1744
      5 A2369
      10 A2898
      20 A37150
      50 A47180
      100 A60300
      1 Specifications indicate the maximum flatness deviation from dc, and include both measured AC-DC difference and the uncertainty of measurement. They are stated at k=2, approximately 95 % confidence.
      2 Includes 1-year stability of the AC-DC difference.
      3 Specifications for the 1 mA current shunt are for TCal ± 1 °C.
      General Specifications
      Calibration Documentation17025 accredited report of calibration included
      Dimensions (maximum)Shunt Value:     1 mA to 2 A
      Height:     70 mm (2.75 in)
      Width:     70 mm (2.75 in)
      Overall Length1:     124 mm (4.9 in)
      Shunt Value:     5 A to 20 A
      Height:     130 mm (5 in)
      Width:     130 mm (5 in)
      Overall Length1:     210 mm (8.25 in))
      Shunt Value:     50 A and 100 A
      Height:     200 mm (7.9 in)
      Width:     200 mm (7.9 in)
      Overall Length1:     343 mm (13.5 in))
      1 Includes input and output connectors; subject to change by component vendor.
      Physical/mechanical parametersShunt Value:     1 mA to 20 A
      Weight:     0.7 kg (1.6 lb)
      Input Connector:     Type-N (female)
      Output Connector:     Type-N (female)
      Shunt Value:     50 A and 100 A
      Weight:     3.4 kg (7.5 lb)
      Input Connector:     Type-LC (female)
      Output Connector:     Type-N (female)
      Operating environmentTemperature:    13 °C to 33 °C
      Calibration temperature (TCal) range:    18 °C to 28 °C
      Humidity range for best specification 1,2:    ≤ 50 % RH
      Altitude:    0 m to 3,000 m
      1 Resistance stability is affected by humidity, but changes are reversible.
      2 If the shunts are calibrated outside this RH, stability specifications will be met as long as the shunts are stored and used at the same relative humidity ± 10 % RH.
      Storage and transit environments (for models other than the 1 mA current shunt)Temperature to avoid damage:    -20 °C to 140 °C
      Temperature and humidity to maintain performance1:    5 °C to 45 °C; 15 % to 80 % RH
      Non-operating altitude:    0 m to 12,000 m
      1 Storage at extremes of temperature or humidity will cause a temporary change of shunt resistance by up to ± 20 ppm. When subsequently stored or used within the limits of the operating environment, the shunts will recover to their original resistance value within 30 days.
      Additional 1 mA current shunt specifications
      Output resistance8 mΩ
      Maximum safe output current11 mA (e.g., 1 V output into 90 Ω)
      Maximum capacitive load800 pF
      Output voltage regulation15 ppm/100 pF
      Maximum output dc v offset± 100 μV (typical ± 25 μV)
      Typical error @ 1 MHz< 2 %
      Battery specifications
      Battery sizeAAA (44.5 mm x 10.5 mm)
      Battery technologyNickel-Metal Hydride (NiMH)
      Number of batteries required8 (in 2 groups of 4)
      Nominal battery voltage1.2 V (4.8 V per group of 4)
      Typical battery capacity800 mAh
      Storage and transit environment to preserve the batteriesLess than 90 days:     -20 °C to 40 °C
      Less than one year:     -20 °C to 30 °C
      Charging time (from fully discharged):     100 minutes
      Maximum operating time between chargesMaximum output load (11 mA):     18 hours
      High impedance load:     24 hours
      Recommended cooling period:     100 minutes
        To prevent loss of battery capacity recharge at least twice per year.

      Models: A40B Series Precision DC and AC Current Shunts

      A40B/SET
      Complete set of shunts including:

      Find a distributor
      • 14 shunts
      • transit case
      • A40B-ADAPT/LC LC Male to LC Male adapter (1)
      • A40B-ADAPT/LCN LC Female to N Male inter-series adapter (1)
      • A40B-LEAD/4MM N to 4 MM double banana connector (2)
      • A40B-LEAD/N N Male to N Male lead (1)
      A40B-001MA
      1 milliamp Current Shunt

      Find a distributor
      A40B-010MA
      10 milliamp Current Shunt

      Find a distributor
      A40B-020MA
      20 milliamp Current Shunt

      Find a distributor
      A40B-050MA
      50 milliamp Current Shunt

      Find a distributor
      A40B-100mA
      100 milliamp Current Shunt

      Find a distributor
      A40B-200mA
      200 milliamp Current Shunt

      Find a distributor
      A40B-500mA
      500 milliamp Current Shunt

      Find a distributor
      A40B-1A
      1 amp Current Shunt

      Find a distributor
      A40B-2A
      2 amp Current Shunt

      Find a distributor
      A40B-5A
      5 amp Current Shunt

      Find a distributor
      A40B-10A
      10 amp Current Shunt

      Find a distributor
      A40B-20A
      20 amp Current Shunt

      Find a distributor
      A40B-50A
      50 amp Current Shunt

      Find a distributor
      A40B-100A
      100 amp Current Shunt

      Find a distributor
      • bvseo_sdk, p_sdk, 3.2.1
      • CLOUD, getContent, 24.04ms
      • QUESTIONS, PRODUCT
      • bvseo-msg: HTTP status code of 404 was returned;
      • bvseo_sdk, p_sdk, 3.2.1
      • CLOUD, getReviews, 22.06ms
      • REVIEWS, PRODUCT
      • bvseo-msg: HTTP status code of 404 was returned;
      Please enable cookies to view product reviews.