      Between 4 and 10 reference modules

      Each Nanoscan system consists of one or more Model 7000 solid-state zener reference modules that plug into a mainframe chassis. The 10-Volt ouputs from these modules are routed via an analog signal backplane to the Nanoscan module where they are averaged to produce an ultra low-noise hardware Average 10 Volt output.

      High sensitivity null detector

      The Nanoscan module also contains a high-sensitivity (0.01ppm resolution) null detector that compares individual 10-Volt reference outputs with the Average output, enabling you to verify the stability of each reference to a very high confidence level. Individual references can also be switched in and out of the hardware average to evaluate their overall contribution.

      Simple integration of external reference standards

      In addition to intercomparing individual internalreferences to the system average, the null detector can also be used to measure external reference standard, providing an easy method of importing or exporting traceability from the Nanoscan system.

      10-Volt Average, Buffered 10-Volt Average and Average 1.018-Volt Outputs

      The Nanoscan module also generates a buffered 10V output to drive resistive loads such as Kelvin Varley dividers, and a 1.018V output for comparison with Weston cells.

      Automatic intercomparison using 7050 volt maintenance software

      Although the Nanoscan can be driven manually via its front panel controls, the full benefits of process automation is achieved when you use it with the 7050 Volt Maintenance Software. This software, which runs under WindowsTM95 or NT, enables you to initiate scan sequences and log all the results into an ExcelTM spreadsheet for later analysis.

      Specifications: 7004N/7010N Nanoscan Volt Maintenance System

      10V Output
      Stability7004N:   90 days: 0.8ppm
       1 year: 1.2ppm
      7010N:   90 days: 0.7ppm
       1 year: 1.0ppm
      Predictability7004N:   ±0.2 ppm/year typicalAfter 5 points, 3 mo. apart
      7010N    ±0.1 ppm/year typical
       After 5 points, 3 months apart
      Temperature Coefficient (15 - 35°C)7004N:   < 0.03 ppm/°
      7010N:   < 0.02 ppm/°
      Noise (rms)7004N:   0.05 to 10Hz: < 0.05 ppm RMS
       Standard deviation of 90 day regression: < 0.06 ppm
      7010N:   0.01 to 10Hz: < 0.03 ppm RMS
       Standard deviation of 90 day regression: < 0.04 ppm
      Hysteresis recovery7004N:   < 0.1 ppm (after conditioning cycle and battery discharge)
      7010N:   < 0.07 ppm
       (after battery discharge)
      Output Resistance7004N/7010N: 500/n Ohm
      Where n = number of modules fitted
      10V 4 wire Output (relative to Average)
      Offset< ±1µV
      Output Current12mA max.
      Load RegulationZero to 2mA:     < 0.1 ppm
      2mA to 12mA:     < 0.5 ppm
      External Standard InputRange:   +9.990 to +10.010V
      Input Impedance:     100 Mohm + 10pF (typic)
      Null Detector
      Range-9999.9 µV to +9999.9 µV
      Measurement ErrorsChan. to Average:    ± 0.3% of difference
      Chan.to out of Average:  ± 0.1% of difference
      External Standard:   ± 0.001%
      Chan. to Chan.:  ± 0.1 µV
      Oven Temperature MonitorAccuracy:    ± 3°C
      Stability/Repeatability:     : ± 0.1°C/year
      Battery
      TypeNiMH
      Backup Period16 hrs from fully charged
      Recharge time (typical)2 hours
      Half Life5 yrs
      Environment
      TemperatureOperating:   +15°C to 35°C
      Transit:     -18°C to 45°C
      Warm-up period:  20 min to ±0.2 ppm 2 hours to final value
      General Specifications
      Power, battery life7004N:   < 6W
      7010N:   < 12W
      12V DC ConnectionYes
      Dimensions7004N:   133 x 449 x 355 mm
       (5.24 x 17.68 x 13.19 inch)
      7010N    265 x 449 x 355 mm
       (10.43 x 17.68 x 13.19 inch)
      Weight7004N:   9.6kg (21.2Ibs)
      7010N:   20kg (44.1Ibs)
      SafetyUL3111, CE marked
      EN61010-1-1:1993/A2:1995
      CETL

      Models: 7004N/7010N Nanoscan Volt Maintenance System

      7004N
      4-Reference 'Nanoscan' Volt Maintenance System

      All models include:
      • Low-Thermal Lead Set
      • Interface Cabling for a second 'Nanoscan'/'Transref' unit
      • Fibre-Optic Cabling for connection to a PC
      • 12V DC Power Supply
      • (requires at least one Model 7000 10-volt Solid State DC Voltage Reference Module)
      7010N
      10-Reference 'Nanoscan' Volt Maintenance System

      • bvseo_sdk, p_sdk, 3.2.1
      • CLOUD, getContent, 67.54ms
      • QUESTIONS, PRODUCT
      • bvseo-msg: HTTP status code of 404 was returned;
      • bvseo_sdk, p_sdk, 3.2.1
      • CLOUD, getReviews, 21.89ms
      • REVIEWS, PRODUCT
      • bvseo-msg: HTTP status code of 404 was returned;
      Please enable cookies to view product reviews.