Fluke 714 熱電偶程控校正器

  • Fluke 714 Thermocouple Calibrator 1
停產

主要功能

  • Measure temperature from TC output
  • Simulate TC output
  • Operable with nine types of thermocouples
  • Calibrate linear TC transmitter with mV source function
  • Selectable °F or °C
  • Thermocouple mini-jack termination
  • Available as accessories: Fluke-700TC1 and TC2 Thermocouple Mini-plug Kits
  • Holster features test lead storage
  • Compatible with toolpack magnetic hanging system
此產品已停產或重新分類

產品概述: Fluke 714 熱電偶程控校正器

The Fluke 714 Thermocouple Calibrator delivers outstanding performance, durability and reliability. The calibrator is compact, lightweight, and easy to carry. With a push button interface similar to the multifunction Fluke 750 Series documenting process calibrators, the 714 is easy to use. It is EMI tolerant, dust and splash resistant, and features a removable battery door for quick battery changes.

規格: Fluke 714 熱電偶程控校正器

Functional Specifications
Measure / simulate thermocoupleRange-200 to 1800°C, depending on type
Resolution0.1°C
Accuracy0.3°C + 10 µV
9 TC types; J K T E R S B per NIST 175 and ITS-90 L U per DIN 43710 and IPTS-68
Measure / simulate mVRange-10 to 75 mV
Resolution0.01 mV
Accuracy0.025% + 1 count
Specifications
DisplayLCD, 5 digit temperature and current sim ultaneous
Environmental Specifications
Operating temperature-10°C to 55°C
Storage temperature-40°C to 60°C
Humidity (without condensation)95% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
35% (50°C to 55°C)
Operating altitude3,000 m max
Safety Specifications
SafetyCSA C22.2 No. 1010.1:1992
Mechanical and General Specifications
Size (L x W x D)187 x 87 x 32 mm
(7.93 x 3.86 x 2.06 in)
WeightApprox. 330 g
(21 oz)
Warranty3 years
Shock and vibrationRandom, 2 G, 5-500 Hz
1 m drop test
Battery typeSingle 9 V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline

機型: Fluke 714 熱電偶程控校正器

Fluke 714
Fluke 714 Thermocouple Calibrator

Thermocouple Calibrator including:

  • Protective Yellow Holster with test lead storage
  • Instructions Sheet (14 languages)
  • single 9V alkaline battery
  • NIST-traceable calibration report and data