5320A-LOAD | High Voltage Load Adapter

  • No image available
停產
此產品已停產或重新分類

產品概述: 5320A-LOAD | High Voltage Load Adapter

Hipot current calibration load resistors

The 5320A-LOAD High Voltage Load Adapter is an accessory to the Fluke 5320A Multifunction Electrical Tester Calibrator and the 5080A High Compliance Multi-Product Calibrator with MegOhm option. It loads a hipot tester to create the leakage current needed to perform hipot tester calibration. It can be used standalone with a digital multimeter and high voltage probe, or with the 5320A built-in multimeter and high voltage probe provided with the 5320A.

The 5320A-LOAD consists of a series of eight power-rated resistors configured to provide eight resistance taps from 10 kΩ to 5 MΩ.

Because of the high voltages created by hipot testers, the 5320A-LOAD is designed to withstand a maximum voltage from 1.2 kV to 5.5 kV, depending on the selected resistor. During the hipot calibration test, the digital multimeter and high voltage probe measures the actual test voltage and leakage current from the hipot tester. The 5320A has a built-in multimeter and comes with a high voltage probe for this purpose. The 5080A does not have a built-in multimeter, and does not come with a separate high voltage probe, so those pieces must be purchased separately.

The 5320A-LOAD comes standard with a traceable calibration certificate including verification data. An optional 17025 accredited calibration certificate is available for an extra charge.

The 8508A/CASE is a sturdy case available for transporting the 5320A-LOAD.

The design of the 5320A-LOAD allows it to be placed on top of or under the calibrator for convenient operation and a small footprint.

規格: 5320A-LOAD | High Voltage Load Adapter

Electrical Specifications
Total resistance range10 kΩ to 5 MΩ
Number of specific resistance values8
Tolerance to Nominal Value10 % (One year, Tcal 5 °C)
Maximum Ratings
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
10 kΩ
1200 V
140 W
Limited to 3 minutes
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
35 kΩ
2000 V
110 W
Limited to 3 minutes
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
50 kΩ
2000 V
80 W
Limited to 3 minutes
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
100 kΩ
5500 V
300 W
Limited to 3 minutes
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
250 kΩ
5500 V
120 W
Limited to 3 minutes
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
500 kΩ
5500 V
60 W
No limit
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
1 MΩ
5500 V
30 W
No limit
Nominal value
Max. voltage
Max. dissipation power
Max. time at maximum power
5 MΩ
5500 V
5 W
No limit
General Specifications
Power supply voltageAC adapter 100-240 V, output voltage 12 V @ 0.4 amps min
Warm-up timeNot applicable
Specifications confidence level99 %
TemperatureOperating Temperature: 5 °C to 40 °C
Recommended Calibration Temperature (Tcal):23 °C
Storage Temperature: -20 °C to +70 °C
Altitude, MaximumOperating: 3,050 m (10,000 ft)
Storage: 12,200 m (40,000 ft)
Dimensions430 mm X 462 mm X 95 mm (16.9 in X 18.2 in X 3.7 in)
Weight (net)3 kg (8 lb 4.5 oz)
Power Consumption5 W maximum
Safety classI according to EN 61010-1