96270A RF Reference Source Firmware Upgrade (.ldr)
Serial Number Range | Backward Compatibility | Date | Firmware Revision | Explanations |
---|---|---|---|---|
All | All previous versions 96270A | April 2, 2019 | V1.13 | Extend capability of previous 96270A models to accept Rohde & Schwarz NRPxxT power sensors. Full list of all sensors recognised by 96270 with install firmware V1.13 installed: Older NRP-Z51, NRP-Z52, NRP-Z55, NRP-Z56, NRP-Z57and NRP-Z58 Now extended to include: New: NRP18T, NRP18TN, NRP33T, NRP33TN, NRP40T, NRP40TN, NRP50T, NRP50TN, NRP67T, NRP67TN and NRP110T. Resolves minor resolved issues |
Firmware Installation Instructions
- Power off the Product.
- On the back panel locate the dip switches and set dip switch 6 to its up position.
- Insert the USB stick.
- Turn on the Product. A yellow screen with an empty progress bar should appear on the screen.
- Follow the on screen instructions.
- When complete, place the dip switch (6) down again to its original position, remove the USB stick and power cycle.