525A Temperature/Pressure Calibrator

No se fabrica

Características principales

Más información
Este producto ya no se fabrica o se reclasificó

Descripción general del producto: 525A Temperature/Pressure Calibrator

At home in the cal lab
The 525A is a compact bench instrument well suited for traditional calibration laboratories with a large temperature workload. Its 0-100V output and 0-100 mA current capabilities – plus its ability to serve as a highly-accurate pressure standard – make it a versatile performer.

Filling the gap in the instrument shop
In process plant environments, instrument shop technicians responsible for maintaining the equipment pool need to calibrate and repair a wide range of instruments, transmitters, field calibrators, P/I converters, transducers, and temperature probes. With its ability to simulate almost any temperature sensor, precise pressure monitoring capability, and dc voltage and current accuracy rivaling any calibrator in its price range, the 525A fills the gap between lower accuracy handheld field calibrators and more expensive high-end multifunction calibrators.

A good fit for ATE test applications
With precise calibration-grade instrument control enabled via an IEEE bus, the 525A is a hard working addition to automated test systems in quality and manufacturing test applications requiring capabilities such as voltage and current sourcing, sensor simulation, and temperature monitoring and validation.

Powerful functionality that is easy to use
From start to finish, the 525A is designed for ease of operation. The intuitive front panel design features large keypads and display that help reduce training time and make the 525A comfortable to use even for long periods of time. Plus, you can store frequently used constants for a variety of probes in memory for faster setup on the job.
MET/CAL® Plus automates 525A calibration and documentation management
You can use the 525A with Fluke’s MET/CAL Plus Calibration Software (v 6.11 or later) to meet the stringent documentation and reporting requirements imposed by quality standards such as ISO 9000. MET/CAL Plus is a powerful software environment for creating, editing, testing, and documenting calibration procedures, and for performing automated calibrations.

Pressure modules
An optional set of external pressure modules provides pressure measurement capability. The 525A can accept either the Fluke 700 Series pressure modules or the Fluke Model 525A-P series pressure modules. These modules plug directly into the calibrator’s front panel Lemo connector; the 525A firmware auto-detects the type and value of the attached module.

Especificaciones: 525A Temperature/Pressure Calibrator

 
Specifications One Year (Source)
Voltage (1 ma Max Load)0 to 100 V  
Best one year specification:     30 ppm of setting
Current (10V Compliance0 - 100 mA  
Best one year specification:     100 ppm of setting
Resistance5 - 4,000 Ohms  
Best one year specification:     0.03 Ohms
Thermocouple (B,C,E,J,K,L,N,R,S,T,U)Standard thermocouple ranges    
Best one year specification:     0.14° C (type E)
Specifications One Year (Measure)
SPRT 25.5 Ohm- 200° C to 660° C  
Best one year specification:     0.02 C
Resistance5 - 4,000 Ohms  
Best one year specification:     0.03 Ohms
Thermocouple (B, C, E, J, K, L, N, R, S, T, U)Standard thermocouple ranges    
Best one year specification:     0.14° C (type E)
RTD*-200° C to 800° C  
Best one year specification:     0.07° C (pt 100)
Ni 120- 80° C to 100° C  
Best one year specification:     0.01° C
Cu 10- 100° C to 260° C  
Best one year specification:     0.06° C
YSI 400 thermistor15° C to 50° C  
Best one year specification:     0.007° C
General Specifications
Settling TimeLess than 5 seconds
Temperature PerformanceOperating:   0° C to 50° C
Calibration (tcal):  15° C to 35° C
Storage:     - 20° C to 70° C
InterfacesTemperature coefficient for temperatures outside tcal + 5° C is 10 % of the 90 day specification per °C
RHOperating:   < 80% to 30° C, < 70 % to 40° C, 40 % to 50° C
Storage:     <95%, noncondensing
Analog Low Isolation20V
EMCIEC 61326- 1/1997 (EMC)
Weight/Dimensions9 lbs (4 kg) ; H 5.25" x W 12.5" x D 18.6" (H 13.1 cm x W 31.3 cm x D 47.3 cm)

Modelos: 525A Temperature/Pressure Calibrator

525A
Calibrador de temperatura/presión

525A-GPIB
Calibrador de temperatura/presión con interfaz GBIB