Estándar de tensión directa 732B

  • No image available
No se fabrica
Este producto ya no se fabrica o se reclasificó

Descripción general del producto: Estándar de tensión directa 732B

Each standard is small, light and rugged, ideal for shipment. The long 72-hour battery life allows the 732B to be shipped over long distances under power. An optional external battery and charger extends battery life still further, to more than 130 hours.

The 732B can stand up to a lot of abuse. The inputs can be shorted indefinitely and are protected up to 1100V dc, 25 mA, without damaging the cell or affecting its output.

Confidence
The 732B is based on the proven technology of the 732A, the first standards lab quality electronic reference to gain acceptance as a replacement for saturated standard cells. Originally designed for internal transfers of the Fluke Calibration corporate volt to the production floor, thousands are now in service world-wide in a variety of applications -- from maintaining an institutional reference to transferring values from national labs or privately-operated 10V Josephson Arrays.

Stability
Stability for each output is ±2 ppm per year (10V) and ±0.8 ppm per month (1.018V). Each 732B can drive up to 12 mA of current to simplify use with instrumentation - like the 5700A - with low output impedance.

Support for artifact calibration
Combined with the 742A-1 and 742A-10k Resistance Standards, a single 732B makes a tough and compact artifact calibration support package for instruments like the 5700A/5720A Calibrators from Fluke Calibration or the 3458A Multimeter from Hewlett-Packard.

Traceability
732Bs can be ordered with or without calibration certificates. The standard 732B comes without any formal traceability or any calibration certificate. Optionally a calibrated 732B can be ordered and requires both shipping in a “power on” condition (there are limits for the overall shipping transportation time). The model 732B/H is a 732B which ships with a Report of Calibration showing the deviation from nominal and the uncertainty at the time of the calibration. The 732B/C is similar to the 732B/H, but it adds a drift rate characterization report. This is a 12 month projection into the future from the time of calibration of the changes in the calibrated output value.

Especificaciones: Estándar de tensión directa 732B

Output
10V30 Day Specification:     ± 0.3 ppm
1 Year Specification:     ± 2.0 ppm
1.018V30 Day Specification:     ± 0.8 ppm
1 Year Specification:     n/a
General Specifications
Temperature15°C to 35°C, operating; -51°C to 71°C, non-operating (with internal battery pack switched off)
Relative Humidity<95% to 30°C, <75% to 35°C, non-condensing
Altitude<3049m (10,000 ft) operating, <12,195m (40,000 ft) non-operating
VibrationPer MIL-T-28800; Type III, Class 5, Style E
SafetyDesigned to IEC 348, 2nd edition; 1978 and ANSI/ISA-S82; and UL1244, 2nd edition 1980, CSA C22.2 No 231, and IEC 1010
Calibration DocumentationReports of Calibration are optional
Size732B DC Standard:   13.4 cm H x 9.8 cm W x 40.6 cm D (5.28 in H x 3.86 in W x 15.98 in D)
732B-7001 External Battery and Charger:   13.4 cm H x 9.8 cm W x 40.6 cm D (5.28 in H x 3.86 in W x 15.98 in D)
Weight732B DC Standard:     5.91 kg (13 lbs.)
732B-7001 External Battery and Charger:     5.45 kg (12 lbs.)
Power, battery life732B DC Standard:     100V, 120V, 220V, 240V ac ± 10%; 50 Hz to 60 Hz; 66W max. Internal 12V lead acid, gelled electrolyte battery operates for 70 hours at 23°C when fully charged. Trickle-charged continually when external ac power is applied.
732B-7001 External Battery and Charger:     100V, 120V, 220V, 240V ac ± 10%; 50 Hz to 60 Hz; 66W max. Internal 12V lead acid, gelled electrolyte battery operates for 70 hours at 23°C when fully charged. Trickle-charged continually when external ac power is applied.

Modelos: Estándar de tensión directa 732B

732B
Patrón de CC. No se incluye la calibración de salida.

732B/C
Patrón con caracterización de deriva y certificado de calibración incluidos, enviado bajo tensión

732B/H
Patrón. Con certificado de calibración incluido, enviado bajo tensión