1575A/1590 Super-Thermometer Readout
No se fabrica
Este producto ya no se fabrica o se reclasificó
Fluke sugiere:Supertermómetros 1594A/1595A
Descripción general del producto: 1575A/1590 Super-Thermometer Readout
Especificaciones: 1575A/1590 Super-Thermometer Readout
Specifications | 1575A | 1590 | ||||
Nominal Resistance | Accuracy (of indicated value) | Equivalent Temp. Value, at 0 °C | Nominal Resistance | Accuracy (of indicated value) | Equivalent Temp. Value, at 0 °C | |
Transfer Accuracy (using external reference resistor) | 0.25W 2.5W 25W 100W 10 KW | 40 ppm 20 ppm 4 ppm 4 ppm 10 ppm | 0.01 °C 0.005 °C 0.001 °C 0.001 °C 0.00025 °C (thermistor at 25 °C) | 0.25W 2.5W 25W 100W 10 KW | 20 ppm 5 ppm 1 ppm 1 ppm 5 ppm | 0.005 °C 0.00125 °C 0.00025 °C 0.00025 °C 0.000125 °C (thermistor at 25 °C) |
Absolute Accuracy (using internal reference resistor) | 0.25W 2.5W 25W 100W 10 KW | 100 ppm 40 ppm 8 ppm 8 ppm 20 ppm | 0.025 °C 0.01 °C 0.002 °C 0.002 °C 0.0005 °C (thermistor at 25 °C) | 0.25W 2.5W 25W 100W 10 KW | 40 ppm 20 ppm 6 ppm 6 ppm 10 ppm | 0.01 °C 0.005 °C 0.0015 °C 0.0015 °C 0.00025 °C (thermistor at 25 °C) |
Typical Resolution | 0.25W 2.5W 25W 100W 10 KW | 10 ppm 5 ppm 1 ppm 1 ppm 3 ppm | 0.0025 °C 0.00125 °C 0.00025 °C 0.00025 °C 0.000075 °C (thermistor at 25 °C) | 0.25W 2.5W 25W 100W 10 KW | 10 ppm 2 ppm 0.5 ppm 0.5 ppm 2 ppm | 0.0025 °C 0.0005 °C 0.000125 °C 0.000125 °C 0.00005 °C (thermistor at 25 °C) |
Resistance Range | 0W to 500 KW | |||||
Internal Reference Resistors | 1W, 10W, 100W, 10 KW | |||||
Minimum Measurement Period | 2 seconds | |||||
Current Source | 0.001 mA to 15 mA, programmable | |||||
Analog Output | –5 to +5 V | |||||
Display | Monochrome LCD with CCFT backlight | Color LCD with CCFT backlight | ||||
Power | 100–125/200–250 VAC (user switchable), 50/60 Hz, 1 A | |||||
Size/Weight | 178 mm H x 516 mm W x 320 mm D (7.0 x 20.3 x 12.6 in) / 16 kg (35 lb.) | |||||
Calibration | Includes NIST-traceable accredited calibration |