Fluke 2600A-101 Sprt Probe

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 198.57ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • Fluke 2600A-101 Sprt Probe

Key Features

  • The 2600A-101 is a precision four-wire, 100 Ohm SPRT (Secondary Platinum Resistance Thermometer) probe that have been designed and tested to meet the precision required as a secondary reference standard.

Product overview: Fluke 2600A-101 Sprt Probe

Secondary Platinum Resistance Thermometer Probe
The 2600A-101 is a precision four-wire, 100 Ohm SPRT (Secondary Platinium Resistance Thermometer) probe that have been designed and tested to meet the precision required as a secondary reference standard. The 2600A-101 has stripped leads suitable for connection to any Fluke data acquisition product or other screw terminal connector. This probe comes with calibration certificates and data for -200 to 6600°C.

Probe hook up diagram:
The probe connects to channel 1(input) and 11(excitation) on the Hydra input connector.

2600A-101 SPRT Probe

2600A-101 SPRT Probe

Warning: Probe handle surfaces are conductive. Use caution near electric circuits. Voltage levels above 30V rms, 42 V peak, or 60 V dc should not be applied to the probe.

Specifications: Fluke 2600A-101 Sprt Probe

General Specifications
Use Temperature200 °C to 660 °C
StabilityTypical draft < 0.01°C after 100 times from 25 to 660 °C
Calibrated Accuracy
TemperatureUncertainty
- 183°C+/- 0.01°C
0°C+/- 0.007°C
200°C+/- 0.024°C
75%°C+/- 0.033°C
660 °C+/- 0.050 °C
Performance Specifications (nominal)
Ice point resistance:100 Ω ± 0.1 Ω
Alpha value:0.00385 ± 0.00001 Ω /Ω /C
Insulation resistance:1000 Meg Ω at 100 Vdc at 20°C
 100 Meg Ohms at 100 Vdc at 400°C
Hysteresis:0.01°C Max. using 0, 400°C as the end points
Self heating:0.02°C / mw in well stirred bath
Thermoelectric voltage:When tested in an ice bath, with immersion depth from 4 to 10 inches (10 to 25 cm), thermo-electric voltage will not exceed 2 µV
Time constant:9 seconds typical for 63.2% response to step change
Long term stability:Stability at 0°C after exposure to upper temperature limits
Dimensions
Sheath diameter (D):.25 inches (6.3 mm)
Length (L):12 inches (0.3 m)
Cable length (Y):6 feet (1.83 m)
Wire:22 AWG, PTFE jacketed, rated to 200 C
Wire terminations:flying lead; stripped / tinned

Models: Fluke 2600A-101 Sprt Probe

2600A-101

SPRT Probe


Buy it
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 25.73ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 26.74ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.