7004T/7010T
Product overview: 7004T/7010T
Simple means of importing traceability
While Fluke’s Nanoscan Volt Maintenance Systems are designed to sit permanently in your standards lab, the Model 7004T Transref provides a simple means of importing traceability from higher order standards laboratories. It can equally be used to export your 10-Volt traceability to other labs.
Safely transportable
Safely stowed in its ruggedized transit case, the Model 7004T allows up to four Model 7000 reference modules to travel, without fear of mechanical damage or undetected exposure to temperature extremes.
During transportation, each reference module remains powered either from its internal battery pack or from a 12-Volt DC supply (e.g. vehicle cigar lighter socket).
The Transref isn’t solely reliant on its batteries or an external DC supply. Even if the batteries become exhausted and the unit is subjected to temperatures down to -5°C, a patented zener reference conditioning technique ensures that each reference will typically return to within ±0.2 ppm of its previous values when the Transref unit is re-powered. A temperature monitor in the transit case alerts you to the maximum and minimum temperatures experienced during shipment
Low-noise low-drift output
The 10-Volt Average output on the Transref provides a low-noise low-drift output that can be used for importing/exporting traceability or as a means of checking the unit’s stability on its return to your standards lab.
Interface directly to the Nanoscan unit
However, if you’re importing or exporting traceability to or from a Fluke Nanoscan system, the Transref can be interfaced directly to the Nanoscan unit so that its reference modules are included in Nanoscan scan measurement sequence. This allows the associated 7050 Volt Maintenance Software to maintain as comprehensive a history on the Transref as it does on your Nanoscan system.
Permanent extension chassis for Nanoscan system
The Model 7004T Transref, or its ten reference equivalent the Model 7010T, can also be used as a permanent extension chassis for a Nanoscan system, increasing the maximum number of internal reference modules it can accommodate to 20.
Specifications: 7004T/7010T
10V Output | |
Stability | 7004T: 90 days: 0.8ppm 1 year: 1.2ppm 7010T: 90 days: 0.7ppm 1 year: 1.0ppm |
Predictability | 7004T: ±0.2 ppm/year typicalAfter 5 points, 3 mo. apart 7010T: ±0.1 ppm/year typical After 5 points, 3 months apart |
Temperature Coefficient (15 - 35°C) | 7004T: < 0.03 ppm/° 7010T: < 0.02 ppm/° |
Noise (rms) | 7004T: 0.05 to 10Hz: < 0.05 ppm RMS Standard deviation of 90 day regression: < 0.06 ppm 7010T: 0.01 to 10Hz: < 0.03 ppm RMS Standard deviation of 90 day regression: < 0.04 ppm |
Hysteresis recovery | 7004T: < 0.1 ppm (after conditioning cycle and battery discharge) 7010T: < 0.07 ppm (after battery discharge) |
Output Resistance | 7004N/7010N: 500/n Ohm Where n = number of modules fitted |
Battery | |
Type | NiMH |
Backup Period | 16 hrs from fully charged |
Recharge time (typical) | 2 hours |
Half Life | 5 yrs |
Environment | |
Temperature | Operating: +15°C to 35°C Transit: -18°C to 45°C Warm-up period: 0 min to ±0.2 ppm 2 hours to final value |
General Specifications | |
Power, battery life | 7004T: < 6W 7010T: < 12W |
12V DC Connection | Yes |
Dimensions | 7004T: 133 x 449 x 355 mm (5.24 x 17.68 x 13.19 inch) 7010T: 265 x 449 x 355 mm (10.43 x 17.68 x 13.19 inch) |
Weight | 7004T: 9.6kg (21.2Ibs) 7010T: 20kg (44.1Ibs) |
Safety | UL3111, CE marked EN61010-1-1:1993/A2:1995 CETL |