7252, 7252i Dual Output Digital Pressure Controllers

  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getAggregateRating, 132.97ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • No image available

Key features

  • Two independent pressure control channels for simultaneous pressure control and wider overall pressure range coverage
  • Unique Quartz Bourdon Tube Pressure Sensor design provides high precision pressure measurement
  • Available in custom Pressure ranges from 5 to 2500 psi
  • Stable Pressure Control of 0.001 % FS.
  • Cover wider pressure ranges with the 7252i with 0.005 % Reading measurement precision

Product overview: 7252, 7252i Dual Output Digital Pressure Controllers

Adding to the flexibility of the Series 7252, two performance models are available: the Model 7252i that offers percent of reading performance further extending the capabilities of a single calibrator, or the Model 7252, that provides percent of full scale (FS) performance for each range.

The Model 7252i provides a precision of 0.005 % of Reading from 25 % to 100 % of each installed range. For pressures below the lower threshold of 25 %, the precision becomes 0.005 % of the 25 % value. Alternatively, the Model 7252 provides a precision of 0.003 % of FS for each installed range. Both models provide a one year stability of 0.0075 % of Reading allowing a one year calibration interval.

Both the 7252i and 7252 are available in a variety of pressure ranges. Choose from any pressure range combination from as low as 10 inH2O (25 mbar) up to 2500 psi (172 bar). For example, either the Model 7252 or 7252i can be configured with a lower pressure range of 10 psi (1 bar) and an upper range of 1 000 psi (70 bar). Or, a Model 7252i could include a 10 inH2O (25 mbar) range and 2 500 psi (172 bar) to cover very low and very high pressure calibration requirements. Only Fluke Calibration provides such a broad pressure range coverage factor in a dual output pressure controller.

For absolute mode calibrations, the Series 7252 can be supplied with a Barometric Reference option, or the Vacuum Reference option where a vacuum pump is connected to the reference port(s). While the former offers a high degree of convenience by allowing instantaneous transition from gauge to absolute mode, the Vacuum Reference option provides the ultimate performance solution, since it does not introduce the additional uncertainty of a secondary, barometric sensor. Instruments with the Vacuum Reference option include an on-board vacuum sensor to allow fully automatic zeroing when operating in absolute mode.

For applications where absolute calibrations are performed exclusively, permanent absolute ranges to 50 psia (4 bar) are available. The Series 7252 also provides negative gauge operation to support the calibration of a wide variety of test devices and applications.

The Series 7252 is a dual output pressure controller, which increases throughput, while providing maximum stability for each range. With simultaneous pressure control, it can actually perform two calibrations at the same time.

For Series 7252s configured with dual ranges between 10 (1 bar) and 1 000 psi (70 bar), the two test ports can be interconnected. The instrument will then auto range from the lowest to the highest sensor and vice versa. For performing single calibrations, this feature provides a high degree of performance over the full range. Increased throughput is also accomplished via the digital control valve technology, that ensures fast control speed with no overshoot. For example, each range will control in 10 % increments in 15 seconds, or less, into a 245 cubic centimeters (15 cubic inch) volume. No other controller combines this speed coupled with no overshoot control. For ease of use, the large, bright, active matrix color display provides the operator with menus that are presented in plain text with a logical navigation structure similar to all other Ruska digital pressure controllers and indicators, with multi-lingual support.

Automating pressure test and calibration

The Model 7252i and Model 7252 are easy to use and can automate your calibrations in several ways:

Step up/down: For calibrations where the increments are fixed intervals, enter a user-defined step value. The Series 7252 increases or decreases the pressure by the step amount with the jog dial—no lengthy keystroke sequences to program.

Sweep test: For simple exercising routines, as with dial gauges, enter a start value, a stop value and number of times to repeat the cycle. The Series 7252 will automatically exercise the device under test prior to the calibration run.

Computer interface: Every Series 7252 is provided with both an RS-232 and IEEE-488 interface, and all Series 7252’s syntax follow SCPI protocol for easy programming. Intecal, an off-the-shelf software package, is available in addition to a LabVIEW® driver, a free download. Firmware updates can also be performed over the RS-232 interface (updates can be downloaded from the website). A MET/CAL driver is also an available option.

Versatility to handle any pneumatic pressure calibration

The Series 7252 is versatile enough to handle almost any type of pneumatic pressure calibration.

Wide pressure range: The Series 7252 is available in a variety of full scale (FS) pressure ranges from 10 inH2O (25 mbar) to 2 500 psi (172 bar).

Pressure units/scales: Select from over fifteen standard units of measure, including inHg at 0 °C and 60 °F, kPa, bar, psi, inH2O at 4 °C, 20 °C and 60 °F, kg/cm2, mmHg at 0 °C, cmHg at 0 °C, and cmH2O at 4 °C, and two user defined units.

Head pressure: The Series 7252 automatically corrects for head pressure differences.

Autovent and autozero: With a few keystrokes, the Series 7252 will vent the test port to atmosphere or automatically zero itself (autovent is not applicable to permanent absolute models).

Protection of the device under test: Set upper and lower pressure limits to ensure protection of the device under test.

Specifications: 7252, 7252i Dual Output Digital Pressure Controllers

Pressure Ranges
Standard Pressure RangesModel 7252i
Select two of any Full Scale pressure range from 5 to 2500 psig (400 mbar to 172 barg), or permanent absolute range from 15 to 50 psia (1 to 4 bara). Low Pressure (LP) ranges starting at 10 inH2O (25 mbarg) are also available. Please refer to the 7250LP data sheet for additional information on LP ranges and specifications.

Model 7252
Select two of any Full Scale pressure range from 5 to 2500 psig (400 mbar to 172 bar). Or, any permanent absolute range from 15 to 50 psia (1 to 4 bara).
Optional ModesAbsolute using barometric reference sensor for ranges from 15 to 2500 psig (1 to 172bar)

Absolute using Vacuum Reference option for ranges from 5 to 2500 psig (400 mbar to 172 bar)
Negative gauge
Performance
PrecisionModel 7252i
Ranges from 5 to 2500 psi (400 mbar to 172 bar) From 25% to 100% FS: 0.005% of reading Below 25% FS: 0.005% of 25% FS

Model 7252
Ranges 5 to 2500 psi (400 mbar to 172 bar) 0.003% of FS

Model 7252i
0.36/1 psi (25/70 mbar) range: 0.005% of reading from 10% to 100% of maximum FS

Precision is defined as the combined effects of linearity, repeatability and hysteresis throughout the operating temperature range.Expression of uncertainty conforms with the recommendations of the ISO Guide to the Expression of Uncertainty in Measurement.
Stability0.0075% of reading per year
Control StabilityActive mode: 0.001% of each range
0.36/1 psi (25/70 mbar) range: 0.004% of each
Range Passive mode: No additional uncertainty
Display ResolutionUser selectable to 1:1,000,000
Control Response15 seconds or less with zero overshoot into a 15 in3 (245 cc) volume in 10% increments
Negative Gauge Precision (Optional)Model 7252i
0.005% of 25% FS or 0.0005 psi* (0.0345 mbar)

Model 7252
0.003% of positive FS

*Whichever is greater
Barometric Reference (Optional)0.002 psi (0.1379 mbar) maximum error per year
Vacuum Reference (Optional)0.0002 psi (0.014 mbar) maximum error per year
Calibration
CalibrationA calibration report with traceability to National Institute of Standards and Technology (NIST) is provided. bar). A NVLAP accredited calibration is available.
Total Uncertainty
Maximum deviation from the true value of pressure including precision, stability, temperature effects and the calibration standardModel 7252i
Ranges 0.36 to 2500 psi (25 mbar to 172 bar) (25%-100%FS)
90 day: 0.006% reading
1 year: 0.009% reading

Series 7252 Ranges 5 to 2500 psi (400 mbar to 172 bar)
90 day: RSS 0.003% FS + 0.002% reading
1 year: RSS 0.003% FS + 0.0075% reading
Control ParametersVolume:
5 to 60 cu.in. (82 to 980 cc)

Control low limit:
0 psi (0 mbar) gauge
0.15 psi (10 mbar) absolute
Communications
CommunicationsRS232 and IEEE-488, SCPI syntax
Firmware updates are performed via RS232 interface
DriversLabVIEW driver
Languages
Menus and FunctionsThe Series 7252 is capable of displaying menus and functions in: English, Japanese, French, Spanish, Chinese, Italian and German
Options
Options
  • Barometric reference (absolute and negative gauge)
  • Vacuum reference* (absolute)
  • Negative gauge only
  • NVLAP accredited calibration
  • Rack mount kit
  • MET/CAL driver
  • Intecal software
  • Liquid trap assembly

*Requires external vacuum pump

General
DisplayTFT, VGA, Active matrix, 6.4 in (163 mm), 640 X 480 resolution, 65,000 colors
TemperatureOperating: 64°F to 97°F (18°C to 36°C)
Storage: -4°F to 158°F (-20°C to 70°C)
Humidity5% to 95% relative humidity, non-condensing
Dimensions (h x w x d)All versions: 7 in X 16.5 in X 19 in (178 mm x 419 mm x 483 mm)
WeightSeries 7252/7252i: 20 lb (9 kg)
Power90 to 260 VAC, 50/60 Hz, 150 W
Test Port and Supply Connection1/4 inch NPT female
Warm up Time2 to 3 hours; may be left on indefinitely
Pressure MediumNitrogen or clean dry air

Models: 7252, 7252i Dual Output Digital Pressure Controllers

7252
Dual Output Digital Pressure Controller

Buy it
With percent of full scale (FS) performance for each range
7252i
Dual Output Digital Pressure Controller

Buy it
With percent of reading performance
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getContent, 120.78ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
  • bvseo_sdk, p_sdk, 3.2.1
  • CLOUD, getReviews, 22.87ms
  • REVIEWS, PRODUCT
  • bvseo-msg: HTTP status code of 404 was returned;
Please enable cookies to view product reviews.